You are here: Home> Products> Syringe Pump
Brief Description

●   Compatible with disposable syringes from 5ml,10ml,20ml,30ml,50ml,60ml.
●   Stackable to be the maximum 8 channels.
●   High accuracy +/-2% after correct calibrated.
●   Three work modes: Rate mode, Time mode and Weight mode.
●   Bolus / Purge, KVO function.
●   With drug libraries:10 categories of common drugs,174 kinds in total.
●   History recorded, real time displayed.
●   6 hours continuous work after fully charged.

Main Technical Parameters  
Applicable syringes :
5ml,10ml,20ml,30ml,50ml,60ml of any standard
Accuracy :
+/- 2% with correct calibration
Rate mode :
50ml(60ml):0.1ml/h-1800.0ml/h (0.1ml/h step)
20ml(30ml):0.1ml/h-800.0ml/h (0.1ml/h step)
10ml:0.1ml/h-400.0ml/h (0.1ml/h step)
Time Mode :
SOL. VOL:0.1-999.9ml (0.1ml step)
TIME:1-1999min (1min step)
Body Weight Mode :
Volume:0.1-999.9ml(0.1ml step)
DRUG MASS:0.1-999.9mg 0.1mg step)
DOSE:0.01-99.99ug/kg/min(0.01ug/kg/min step)
WEIGHT:0.1-300kg(0.1kg the least step)
Total VolumeRange :
0.1-1999.9ml (0.1 ml step)
Purge Rate :
50ml(60ml):1800.0ml/h,
20ml(30ml):800.0ml/h,
10ml:400.0ml/h, 5ml: 200ml/h
KVO Rate :
1ml/h
Occlusion pressure :
Three level setting LOW, MEDIUM or HIGH
Alarms :
Near empty, Syringe empty, Volume complete (when limit volume is set), Occlusion, Syringe loose, Operation error, Low battery, device failures.
Three levels of alarm volume LOW, MEDIUM or HIGH
Bolus Rate :
10ml:10-300ml/h,20ml(30ml):10-600ml/h,
50ml(60ml):10-1200ml/h
Bolus Volume :
Maximum of 25ml
Interface :
RS-232 (Reserved port)
Power source :
AC power supply: AC 100 to 240V( +/-10%),50/60Hz;
Built-in Lithium Battery:  Rechargeable, 11.1V, 2000mAh.
8 hours of continuous operation.
Accessories :
AC power line, Operational manual
Protection Class :
Class I, Type CF,IPX4
Dimension :
Net weight :
230mm*150mm*125mm
2.2kg
 