You are here: Home> Products> Vital Sign Monitor
Brief Description
• Suitable for adult, pediatric and neonatal patient
• Standard parameters: NIBP, Pulse Rate and SpO2
• Built-in memory, store up to 100 hours of tabular of patient data and 1000 Pieces of NIBP record
• High brightness cold cathode fluorescent backlight LCD display and LED display
• Advanced digital processing technology providing accurate measurement during low perfusion and movement situation
• Adjustable audible and visual alarms
• Built-in large capacity rechargeable lithium battery support up to 10 hours working-time

Main Technical Parameters

Physical Specifications
Dimension and Weight
Dimension
230mm(W)*90mm(H)*280mm(L)
Weight
standard configuration: 2 kg(with battery)
Operation Environment
Power Needs
AC 100-250V, 50/60Hz,
Power
20VA
Temperature
5-40
Humidity
15-93% non-condensing
Patient range    
Neonate, pediatric, and adult patients
Safety
Meet the requirement of IEC60601 series
Performance Specification
Display                             
LCD + LED
Display mode
1 Plethysmogram waveform
Interface
 
Parameter cable interface
AC power input socket
Battery
Lithium battery inside, 3hours for charging,6hours for continuous working
Trend memory
96 hours
Alarm Indicator
Alarm indicator light, power indicator light pulse tone alarm sound button tones
Alarm
 
User-adjustable high and low limits
3-level audible and visual alarm
NIBP
Method
Automatic oscillometric
Operation Mode
Manual/Automatic/STAT
Auto measure time
Adjustable
Measurement Unit
mmHg
Test interval
1-240minutes,Adjustable
Measurement range
 
 
Adult: 10-270 mmHg ;    
Child: 10-240 mmHg ;     
Infant: 10-135 mmHg
Over-pressure protection
Double safety protection
Accuracy of blood pressure      measurement
The mean error shall be less than±5mmHg
the standard deviation shall be less than 8mmHG
Resolution
1 mmHg
Measuring Interval in AUTO Mode
1,2,3,4,5,10,15,30,60,90,120,180,240 min
Measuring Period in STAT mode
5 min
SpO2
Measurement range
0%-100%
Resolution
1%
Accuracy
±2% (70-100%, Adult/Pediatric, non-motion )
±3% (70-100%, Neonate, non-motion )
±3% (70-100%, Adult/Pediatric/Neonate, motion )
0-69% unspecified
Alarm range
Alarm high limit range 31-100%
Alarm low limit range 20-99%
Refreshing rate
0.5s
Pulse rate
Measurement range
20-250bpm
Resolution
1bpm
Accuracy
2%or2bpm
Alarm
Alarm high limit range:35-250bpm
Alarm low limit range:30-245bpm
Refreshing rate
0.5s
Pulse strength
8-seg bar graph